| Model | PS2050 spectrocolorimeter |
| Optical Geometry | D/8(diffused illumination, 8-degree viewing angle), SCI/SCE (specular component included/specular component excluded) Mode, Comply to CIE No.15,GB/T 3978,GB 2893,GB/T 18833,ISO7724-1,ASTM E1164,DIN5033 Teil7 |
| Characteristic | Adopt CMOS dual beam splitting sensor;Used for color difference quality control in plastic electronics, paint and ink, textile and garment printing and dyeing, printing, ceramics and other industries, also for fluorescence sample measurement. |
| Light Source | Combined full spectrum LED light source, UV light source |
| Integrating Sphere Size | Φ40mm |
| Sensor | CMOS dual beam splitting sensor |
| Wavelength Range | 400-700nm |
| Measuring Aperture | Two Apertures:MAV:Φ8mm/Φ10mm;SAV:Φ4mm/Φ5mm |
| Specular Component | SCI/SCE |
| Color Space | CIE LAB,XYZ,Yxy,LCh,CIE LUV,s-RGB,HunterLab,βxy,DIN Lab99 |
| Color Difference Formula | ΔE*ab,ΔE*uv,ΔE*94,ΔE*cmc(2:1),ΔE*cmc(1:1),ΔE*00, DINΔE99,ΔE(Hunter) |
| Other Colorimetric Index | Spectral reflectance, WI(ASTM E313,CIE/ISO,AATCC,Hunter), YI(ASTM D1925,ASTM 313),Metamerism Index Mt,Staining Fastness, Color Fastness,Color Strength, Opacity,555 tone classification, Munsell(C/2)(Partially realize through the PC software /APP software) |
| Observer Angle | 2°/10° |
| Illuminant | D65,A,C,D50,D55,D75 ,F1,F2(CWF),F3,F4,F5, F6,F7(DLF),F8,F9,F10(TPL5), F11(TL84),F12(TL83/U30), U35,NBF,ID50,ID65(Partially realize through the PC software /APP software) |
| Display | Spectrogram/Values, Samples Chromaticity Values, Color Difference Values/Graph, PASS/FAIL Result, Color Simulation, Color Offset |
| Measuring Time | About 1s |
| Repeatability | Chromaticity value: MAV/SCI, within ΔE*ab 0.03 (After preheating and correction, the average value of the whiteboard was measured for 30 times at an interval of 5s);Spectral reflectance: MAV/SCI, Standard deviation within 0.08% (400nm to 700nm: within 0.18%) |
| Inter-instrument Error | MAV/SCI, Within ΔE*ab 0.3 (Average for 12 BCRA Series II color tiles) |
| Displayed Accuracy | 0.01 |
| Measured Reflectance Range | 0-200% |
| Reflectance Resolution | 0.01% |
| Measurement Mode | Single Measurement, Average Measurement(2-99times) |
| Locating Method | Stabilizer position+camera locating |
| White Calibration mode | Contact type automatic calibration |
| Dimension | 94X68X188mm |
| Weight | About 410g |
| Battery | Lithium battery,3.7V,5000mAh,8000 cycles in 8 hours |
| Illuminant Life Span | More than 1.2 million measurements over 10 years |
| Screen | 2.8 inch TFT true color, Capacitive Touch Screen |
| Interface | USB,Bluetooth®5.0 |
| Data Storage | Standard 500Pcs, Sample 10000Pcs(One data is able to include SCI/SCE); PC mass storage |
| Software support | Andriod,IOS,Windows,Wechat small program, Hongmeng |
| Language | Simplified Chinese, English, Traditional Chinese |
| Operating Environment | 0~40℃, 0~85%RH (no condensing),Altitude < 2000m |
| Storage Environment | -20~50℃, 0~85%RH (no condensing) |
| Standard Accessory | Power Adapter, USB Cable, User Guide, SQCX PC Software(Download from office website), White and Black Calibration Cavity, Protective Cover, Wrist strap, Aperture |
| Optional Accessory | USB Micro Printer, Powder Test Box |
| Notes | Technical parameters are only for reference, subject to the actual sale of the product |
Analytical Instruments
Spectrocolorimeter PS2050
"Fat girl" is the nickname of PS20 series domestic color spectrometer. It is serious about color measurement and high user satisfaction.
PS2050 "Fat girl" spectrocolorimeter is 3nh developed by CMOS dual-channel spectrocolorimeter sensor, which is not only a colorimeter, but also a spectrocolorimeter,with excellent repeatability and the inner-instrument error. The measurement data is stable, accurate and reliable. with a comfortable and round appearance more people love it. It also has quality management software facilitate quality monitoring and data management.
PS2050 spectrocolorimeter is equipped with double apertures MAV:Φ8mm/Φ10mm、SAV: Φ4mm/Φ5mm flat , widely used for plastic electronics, paint ink, textile and garment printing, printing, ceramics and other industries color difference quality control. UV is included for the measurement of fluorescent samples.
Technical Advantages
1、 Adopt D/8 optical structure and SCI/SCE mode.
PS2050 spectrocolorimeter adopts D/8(diffused illumination, 8-degree viewing angle) which is widely applicable in the world, and SCI/SCE (specular component included/specular component excluded) Synthesis technology. It is suitable for color management and quality control in various industries such as color matching and coating, textile, plastic, food, building materials, cosmetics, etc.
PS2010 spectrocolorimeter is equipped with single aperture:Φ8mm flat, widely used for plastic electronics, paint ink, textile and garment printing, printing, ceramics and other industries color difference quality control.
Technical Advantages
1、 Adopt D/8 optical structure and SCI/SCE mode.
PS2010 spectrocolorimeter adopts D/8(diffused illumination, 8-degree viewing angle) which is widely applicable in the world, and SCI/SCE (specular component included/specular component excluded) Synthesis technology. It is suitable for color management and quality control in various industries such as color matching and coating, textile, plastic, food, building materials, cosmetics, etc.
✓ Requirement-Based Sourcing▤ Documentation Support◎ Technical Assistance
Product Documents
Datasheets, manuals, calibration documents, COA/SDS or related technical documents are available where applicable to the selected product and manufacturer.
Need a specific document? Add this product to your RFQ Basket and mention the required document in the item notes.